| 719al Specifications |
| Standard |
| Design Type: |
Variable Volume, XYZ movement, syringe loading |
| Tray Capacities: |
50 (15mm), 72 (12mm), or 128 (8mm) sample vials. Also, four 12 mm vials for standards or rinses |
| Injections Per Vial: |
Unlimited |
| Rinse Strokes |
0-99 for needle port and needle |
| Precision: |
Less than or equal to ±0.5% RSD, for 10µL partial loop injections under ACI test conditions |
| Linearity: |
r2 <= 0.99 |
| Carryover: |
Less than 0.01% with rinse under ACI test conditions |
| Sample Waste: |
None (or user selectable) |
| Keypad: |
20 key, 4 line, 80 character LCD |
| Injection Parameters: |
Partial or fixed loop, sample draw speed, delay time, needle depth, pre-injection volume, pre/post-air volume, rinse strokes, rinse rate, pre pierce |
| Syringe: |
50, 250 (standard), or 2000 µL total volume |
| Valve: |
Valco Cheminert C3 valve with 300µL loop and high speed actuator |
| Programming: |
10 Methods, maximum 128 steps per method. Priority sampling |
| Communications: |
Interfaces: RS-232-C, IEEE, I2C Outputs:BCD, parallel printer port, inject market, 4 user selectable NO or NC Contact Closures with user selectable closure duration times. |
| Dimensions: |
33cm H x 32cm W x 55cm D (13" x 13" x 21") Weight: 14 Kg (31 lbs.) Weight: 17 Kg (38 lbs.) Refrigerated |
| Wetted Materials: |
Teflon, 316 stainless steel and nitronic 60 |
| Voltage: |
Autoselectable: 100-240 VAC ±10%, 47-63 Hz |
| Options |
| Syringe & Trays: |
50µL and 2mL volumes 128 sample vials (8mm), will work with WISP vials 50 sample vials (15mm) Adapter for microplate trays |
| Valves: |
C2 or C4 internal loop Cheminert Valves available |
| Refrigeration: |
4º C to 40º C within ±1° C accuracy using Peltier cooling of entire tray surface |
| Materials: |
Biocompatible materials available |
| Other: |
Valveless models available for FIA and other direct injection applications |
| Accessories |
| AutoChanger 31: |
High Throughput microplate or vial tray changer that interfaces with the Model 719al autosampler to allow unattended sampling for up to 31 microplates. |
| Column Switcher: |
Column compartment designed to hold up to 6 columns. User selectable six position Valco valve can be controlled either manually or via RS-232 connection. Standard and heated versions available |
| Printer: |
Small footprint thermal printer for recording autosampler methods. |
| 3 Channel Degasser: |
Three channel On-Line degassing system for minimizing residual dissolved gases during continuous operation. |
| Product Part Number |
| 719/00000/01 |
Standard Model 719al Variable Volume Autosampler |
| 719/00000/01R |
Refrigerated Model 719al Variable Volume Autosampler |
| 708/25921/00 |
Optional UV Cover. (recommended to be purchased with autosampler) |
| 719/20801/00 |
Software upgrade for performing pre-column derivatization, linking autosampler methods, and with the addition of the Column Sitching Unit, selecting column temperature and one of six columns to be switched into the chromatograph flow path. |
| 719/20802/00 |
Alcott Windows Control Software |